Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography

Open access0 citations

Abstract

Purpose: To evaluate the inter-device agreement between hyperparallel optical coherence tomography (HP-OCT) and swept-source OCT (SS-OCT) in quantitative anterior segment parameters. Methods: Phakic subjects aged ≥40 years were consecutively recruited at a tertiary eye hospital. Horizontal B-scans (0°–180°) were analyzed. On SS-OCT, parameters were measured automatically; on HP-OCT, they were extracted manually using 3D Slicer. Agreement was evaluated using intraclass correlation coefficients (ICCs), Bland–Altman plots, and orthogonal regression. Results: A total of 118 sectors from 59 subjects were analyzed. ICCs were ≥0.88 for all parameters except iris thickness (IT) at 750 µm (IT750, ICC = 0.66). HP-OCT consistently measured lower values than SS-OCT for angle opening distance (AOD) at 500 µm (AOD500: –0.024 [–0.036 to –0.013]) and 750 µm (AOD750: –0.032 [–0.050 to –0.015]), trabecular–iris space area (TISA) at 500 µm (TISA500: –0.008 [–0.013 to –0.003]) and 750 µm (TISA750: –0.012 [–0.020 to –0.004]), iris curvature (IC: –0.015 [–0.021 to –0.009]), IT750 (–0.031 [–0.041 to –0.022]), and anterior chamber depth (–0.118 [–0.138 to –0.098]). For the following parameters, differences between devices increased with larger measurements, with HP-OCT yielding proportionally lower values (proportional bias): AOD500, AOD750, TISA500, TISA750, IC, and lens vault. Conclusions: For anterior segment biometry, HP-OCT and SS-OCT showed high ICCs but limited interchangeability because of systematic and proportional bias. Translational Relevance: Understanding how HP-OCT measurements compare with those from SS-OCT is essential for clinical integration, because inter-device agreement and interchangeability guide how HP-OCT can be applied and interpreted in practice.

// Source

View paper (DOI)Open access versionOpenAlexTranslational Vision Science & TechnologyPublished 2026-10-15

Authors: Andrea Servillo, Natalia Porporato, Bingyao Tan, Benjamin Y. Xu, Tin A. Tun, Shamira Perera, Rahat Husain, Mani Baskaran, Rehena Sultana, Tin Aung, Leopold F. Schmetterer

Institutions: University of Southern California, University of Naples Federico II, Medical University of Vienna, Nanyang Technological University, Duke-NUS Medical School, Singapore National Eye Center, Singapore Eye Research Institute, Sankara Nethralaya