Evaluation of clinical performance of artificial intelligence-assisted measurement of thoracic aortic diameter on non-contrast CT
Abstract
This study aimed to evaluate the clinical performance of artificial intelligence (AI)-assisted measurement of thoracic aortic diameter on non-contrast computed tomography (NCCT). This single-center prospective study included 79 patients (mean age = 70 ± 13 standard deviation; 46 men) who underwent NCCT and arterial phase contrast-enhanced computed tomography (CECT) for evaluating aortic diseases between 2021 and 2022. Maximum diameters of 9 American Heart Association landmarks and the maximum ascending/descending aorta were measured. Upper limits of the 95% confidence interval (CI) for the absolute difference in aortic diameter between the AI-assisted/manual method on NCCT and the manual method on CECT (Diff AI-assisted /Diff manual ) were calculated and compared using a predefined cutoff of 1.5 mm. Evaluation times were also compared. No significant difference was found between per-patient Diff AI-assisted and Diff manual (1.10 mm; 95% CI = 1.03–1.18 vs 1.03 mm; 95% CI = 0.95–1.11, P = .058); the upper limit of the 95% CI was <1.5 mm. The upper limit of the 95% CI of the geometric mean Diff AI-assisted and Diff manual was also <1.5 mm at each landmark. The median evaluation time of the AI-assisted method was shorter than that of the manual method (expert radiologist = 272 seconds [interquartile range (IQR) = 229–313] vs 496 seconds [IQR = 438–550]; P < .001, training radiologist: 304 seconds [IQR = 257–404] vs 710 seconds [IQR = 563–855]; P < .001). AI-assisted thoracic aortic diameter measurement on NCCT was as accurate as manual measurement on CECT; its evaluation time was shorter than that of manual measurement on NCCT.
// Source
Authors: A Suzuki, Nobuo Tomizawa, Kotaro Yamamoto, Yuichi Morita, Shohei Fujita, Katsuhiro Sano, Toshiaki Akashi, Mitsuo Nishizawa, Kazuhiro Suzuki, Akihiko Wada, Daisuke Endo, Satoshi Matsushita, Kanako K. Kumamaru, Keiichi Funaki, Jonathan Sperl, Minoru Tabata, Shigeki Aoki
Institutions: The University of Tokyo, Juntendo University, Massachusetts General Hospital, Kokura Memorial Hospital, Athinoula A. Martinos Center for Biomedical Imaging, Siemens Healthineers (Germany), Siemens Healthcare (United States)