Biologyarticle2026-09-16

Impedance-based tracking of a cochlear implant electrode reliably detects successful array insertion

Open access0 citations

Abstract

Abstract Introduction Continuous impedance measurements during cochlear implantation have been hypothesized to provide a reliable indication if the electrode is inside or outside the cochlea. Therefore they provide a valuable feedback mechanism for successful insertion to the surgeon. This study aims to investigate whether impedances can be used to reliably detect successful insertion of the cochlear -implant electrode array. Methods In a prospective cross-sectional single-center study patients undergoing cochlear implantation with flexible lateral wall arrays (MED-EL, Innsbruck, Austria) were recruited between June 2022 and May 2023. Continuous impedance telemetry during electrode array insertion was obtained and compared to the ground truth of visual and audiological feedback by the surgeon. Results 34 patients (23 women, 11 men;, 16 left ears, 18 right ears) with a mean age of 52.8 years (SD 19.6 years) were included. The present study found a reliability of 88.9% comparing the algorithm of continuous insertion to the ground truth. Conclusions Continuous impedance measurements show a high reliability to detect the insertion of electrode contacts of a cochlear implant electrode array. The data collected by the software are suitable to detect misplacements already during insertion. This method is a promising approach to intraoperatively monitor successful insertion and offers the opportunity to replace fluoroscopy and radiological imaging in the aftercare of cochlear implant surgery.

// Source

View paper (DOI)Open access versionOpenAlexEuropean Archives of Oto-Rhino-LaryngologyPublished 2026-09-16

Authors: Tabita Breitsprecher, Stefan Hans, Aline Föger, Angelika Dierker, Joachim Schmutzhard, Stefan Dazert, Nora M. Weiss

Institutions: Universität Hamburg, University Medical Center Hamburg-Eppendorf, Innsbruck Medical University, Ruhr University Bochum, TUM Klinikum, MED-EL (Austria), St. Elisabeth-Hospital Bochum, Erbe Elektromedizin (Germany)