Technical note: Design, construction, automation, and first-principles calibration of a low volume He measurement line
Abstract
Abstract. A noble-gas analysis line capable of accurate and precise measurements of small absolute amounts of 4He released from crystals is a key analytical step in the production of (U-Th)/He chronologic data. He analysis lines that are custom-built in-house can be optimized for specific lab needs and facilitate continued maintenance, repair, and upgrades. However, there is little information in the published literature about the methods and approaches for building a He line. Here, we describe the design, construction, automation, and metrological calibration of a custom 4He extraction and analysis line as part of establishing laser-ablation (U-Th)/He methods in the University of Colorado Thermochronology Research and Instrumentation Lab (CU TRaIL). The line, called the Jimbochron, is designed to precisely measure very small (∼ fmol) amounts of 4He while being fully automated and easily modifiable in the future. These goals are achieved by minimizing the line volume, adopting a unique double-hexagonal manifold configuration, installing a high-sensitivity quadrupole mass spectrometer, and developing editable LabView code for instrument communication and automation that is straightforward to update. We also explain the steps used to calibrate the Jimbochron metrologically from first principles with a new in-house calibration volume to ensure high-accuracy He measurements. The Jimbochron accurately and precisely measures small gas amounts (< 0.1 fmol) and now routinely generates accurate and precise He data for laser-ablation (U-Th)/He applications.
// Source
Institutions: University of Colorado Boulder